• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES 12X100MM KII OPTICAL ADV FIXATION SYS 6/BOX; GCJ

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES 12X100MM KII OPTICAL ADV FIXATION SYS 6/BOX; GCJ Back to Search Results
Model Number CFR73
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/18/2014
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event product was not returned to applied medical for evaluation.Upon inspection, engineering noted fragmentation of an internal rubber component of the seal.The likely root cause of the damage to the seal is an instrument.Applied medical's instructions for use states that "extra care should be used when inserting angular and asymmetrical instruments, such as "j" hooks and clip appliers.All instruments should be centered axially when inserted through the seal to prevent tearing." applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.This report is being filed as a result of a re-review of applied medical complaints received between (b)(6) 2014 and (b)(6) 2016.This retrospective review was associated with a quality management system (qms) compliance action plan developed and presented to fda to address an (b)(6) 2015 warning letter.Applied medical has revised its mdr reporting criteria to be more conservative and has improved complaint handling and mdr reporting processes.The reviews ensured that recent reportable events were appropriately identified and reported to the designated regulatory authority(ies).This report, which represents the initial and final reports combined, is being submitted based on the findings of that retrospective review.In accordance with 21 cfr 803.56, if additional information is obtained which was not known or was not available when this report was submitted, then a supplemental report will be submitted to the fda.
 
Event or Problem Description
Laparoscopic hepatectomy - "this is the complaint from the market.Physician noticed debris inside patient body four hours later from the beginning of a procedure.The debris was retrieved.Cusa excel probe and ultrasonic probe were used through this unit.Patient status: unknown aomori olmypus received the unit and a piece of black color debris.The debris was completely fit the valve inside the seal.Please analyze this complaint phenomenon and review the device history record of the event unit.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12X100MM KII OPTICAL ADV FIXATION SYS 6/BOX
Common Device Name
GCJ
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6373020
Report Number2027111-2017-01084
Device Sequence Number19024718
Product Code GCJ
UDI-Device Identifier00607915119898
UDI-Public(01)00607915119898(17)170121(30)01(10)1213783
Combination Product (Y/N)N
PMA/510(K) Number
K083638
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/21/2017
Device Model NumberCFR73
Device Catalogue Number101256001
Device Lot Number1213783
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2014
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2014
Initial Report FDA Received Date03/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-