|
Halyard received a single report that referenced two different incidences, which were associated with separate kits, involving the same patient during the same procedure.This is the first of two reports.Refer to 9611594-2017-00033 for the second report.It was reported that during a gastronomy tube placement, all of the sutures placed from the first kit broke during placement there was some stomach discharge during the procedure due to dermal perforation.Another kit was used.No further information has been provided at this time.
|