| Model Number |
DP8730S |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/02/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product has been requested for evaluation; however, it has not been received.Sterilization and lot history records were reviewed and no exceptions were found.Report 1 of 16.
|
| |
|
Event or Problem Description
|
|
Fine metal dust/glittering particles were discovered within the incision.No patient impact or injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Eighteen opened dp8730s pouches from lot v7941 were received in a cardboard box.A total of 18 needle tips were returned.Eight of the needle tips were lying loose in the cardboard box.All 18 were dirty with particulates and dried solutions visible all over.Microscopic examination found that the needles were damaged.The corners of the hub have marring, scratches and gouges with material missing.The damage to the hub is consistent with damage seen when the needle is over-tightened.There was also damage to the tips and scratches on the shafts of the needles.This could contribute to particulate generation during use.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the results of our root cause investigation, it has been determined that the reported particulate is not related to any damage or malfunction of the product.Our investigation has concluded that the observed particles are originating from the metallic cap assembled to the needle wrench which when packaged in the same pouch with the silicone irrigation sleeve results in microscopic particles being transferred from the needle wrench cap to the irrigation sleeve.All products have been found to comply with specifications and function as intended.Additionally, there have been no reports of post-surgical adverse reaction.However, due to increased sightings of microscopic particles and our ongoing continuous improvement activities, we have taken measures to identify an internal cleaning process related to needle wrench assembly to address the reported events.Validation activities have been completed and the process has been implemented.In addition, we are reviewing our component specifications related to particulate matter and will revise specifications, where necessary, to better align with customer expectations.As a result of this additional information, it has been determined that the issue documented in this report does not meet the criteria of a reportable malfunction and the complaint record will be revised accordingly.
|
| |
|