| Catalog Number |
5MAXACE068KIT |
| Medical Device Problem Codes |
Kinked (1339); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/31/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the penumbra system ace 68 reperfusion catheter (ace 68) was kinked approximately 48.5 and 59.0 cm from the hub.Conclusions: evaluation of the returned device confirmed that the ace 68 was kinked.This type of damage typically occurs due to improper handling during removal from the packaging.If the ace 68 is withdrawn from its packaging shell before removing the tubing tray, damage such as this may occur.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|
Event or Problem Description
|
|
During preparation for a thrombectomy procedure, the physician noticed that the penumbra system ace 68 reperfusion catheter (ace 68) was kinked when the technologist handed it to him.The kinked ace 68 was noticed prior to use and therefore, was not used in the procedure.The procedure was completed using a new ace 68.
|
| |
|