• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB MICROFLOW 2.2 NEEDLE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB MICROFLOW 2.2 NEEDLE; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number DP8730S
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/02/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product has been requested for evaluation however it has not been received.Sterilization and lot history records were reviewed and no exceptions were found.Report 7 of 10.
 
Event or Problem Description
Fine metal dust/glittering particles were discovered within the incision.No patient impact or injury was reported.
 
Additional Manufacturer Narrative
Based on the results of our root cause investigation, it has been determined that the reported particulate is not related to any damage or malfunction of the product.Our investigation has concluded that the observed particles are originating from the metallic cap assembled to the needle wrench which when packaged in the same pouch with the silicone irrigation sleeve results in microscopic particles being transferred from the needle wrench cap to the irrigation sleeve.All products have been found to comply with specifications and function as intended.Additionally, there have been no reports of post-surgical adverse reaction.However, due to increased sightings of microscopic particles and our ongoing continuous improvement activities, we have taken measures to identify an internal cleaning process related to needle wrench assembly to address the reported events.Validation activities have been completed and the process has been implemented.In addition, we are reviewing our component specifications related to particulate matter and will revise specifications, where necessary, to better align with customer expectations.As a result of this additional information, it has been determined that the issue documented in this report does not meet the criteria of a reportable malfunction and the complaint record will be revised accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROFLOW 2.2 NEEDLE
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
BAUSCH + LOMB
rochester NY 14609
MDR Report Key6373875
Report Number0001920664-2017-00102
Device Sequence Number1008116
Product Code HQC
UDI-Device Identifier10757770057259
UDI-Public(01)10757770057259(17)210701
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K063331
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 02/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2021
Device Model NumberDP8730S
Device Lot NumberV7469
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/02/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/02/2017
Supplement Report FDA Received Date06/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-