Model Number 5
Medical Device Problem Code Failure To Adhere Or Bond (1031)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/18/2014
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
On march 18, 2016, applied medical issued a voluntary class ii recall of the (b)(4) direct drive clip applier due to increased customer feedback indicating inconsistent clip application, which has the potential to lead to unoccluded vessels.An internal corrective and preventative action (capa) has been initiated to conduct a thorough investigation and implement appropriate corrective actions.Fda has issued recall number z-1621-2016 for this recall.This report is being filed as a result of a re-review of applied medical complaints received between june 1, 2014 and may 31, 2016.This retrospective review was associated with a quality management system (qms) compliance action plan developed and presented to fda to address an april 10, 2015 warning letter.Applied medical has revised its mdr reporting criteria to be more conservative and has improved complaint handling and mdr reporting processes.The reviews ensured that recent reportable events were appropriately identified and reported to the designated regulatory authority(ies).This report, which represents the initial and final reports combined, is being submitted based on the findings of that retrospective review.In accordance to 21 cfr 803.56, if we obtain additional information, which was not known or was not available when this report was submitted, then a supplemental report will be submitted to the fda.
 
Event or Problem Description
Laparoscopic gastric banding- "would not close completely on 5 clips." patient status - unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CA500, EPIX UNIVERSAL CLIP APPLIER 3/BX
Common Device Name
FZP
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6374955
Report Number2027111-2017-00961
Device Sequence Number19025914
Product Code FZP
UDI-Device Identifier00607915117306
UDI-Public(01)00607915117306(17)170807(30)01(10)1229106
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
5
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/07/2017
Device Model Number5
Device Catalogue Number5
Device Lot Number1229106
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2014
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2014
Initial Report FDA Received Date03/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No