Catalog Number 75446540
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.X-ray review: post-op x-rays for l4-s1 tlif shows fractured sacral screw after a fall 4.5 years after surgery.There appears to be a paucity of bone in the l5-s1 disc space.Failure of bony fusion would contribute to this.Ct of the lumbar spine would be useful to see if bony fusion was complete.
 
Event or Problem Description
Pre-op diagnosis: previous l4-s1 tlif procedure: l4-s1 transforaminal lumbar interbody fusion(tlif) levels involved: l4-s1 it was reported that on an unknown date, post-op, patient reported a fall to the surgeon with pain.X-ray conducted on (b)(6) 2016 showed two broken screws.Hence, screw was explanted in a revision l4-s1 fusion surgery.
 
Additional Manufacturer Narrative
Product analysis: visual review confirms screw breakage approximately 1 thread down from the base of the bone screw head.Optic al and microscopic examination of the area of fracture initiation did not identify a pre-existing surface defect near the area of crack origination that could contribute to crack propagation.Fracture surface damage is noted.Microscopic examination of the fracture surface identified a fairly flat fracture surface with multiple ratchet marks near the area of crack propagation, and gently convex progressive striations through the cross-sectional area of the bone screw, consistent with cyclic fatigue.Dimensional examination of the major diameter of the bone screw confirms relevant bone screw dimensions are within print specification.The above observations are consistent with cyclic fatigue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key6375599
Report Number1030489-2017-00456
Device Sequence Number500563
Product Code KWP
UDI-Device Identifier00885074134116
UDI-Public00885074134116
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K042025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number75446540
Device Lot NumberH11L3055
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2017
Initial Date Received by Manufacturer 02/06/2017
Supplement Date Received by Manufacturer07/27/2017
07/27/2017
Initial Report FDA Received Date03/03/2017
Supplement Report FDA Received Date08/10/2017
09/06/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age68 YR
Patient Weight62
-
-