SMITHS MEDICAL ASD, INC. PORTEX® ADULT DISPOSABLE ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C37101307J
Device Problem Air Leak (1008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/01/2017
Event Type  malfunction  
Manufacturer Narrative
Event date: early (b)(6) 2017.Smiths medical has received the sample device.A full evaluation is anticipated, but not yet begun as the device is currently in transit to the investigation site.
 
Event Description
It was reported that a portex® adult disposable anesthesia breathing circuit had an air leak immediately upon use of the device.The device stopped being used due to the incident.It was unknown if any medical treatment was conducted.No patient injury was reported.The event was considered resolved.
 
Manufacturer Narrative
Upon receipt of the returned product specimen, it was visually examined.One disposable anesthesia breathing circuit and breathing bag was returned for evaluation.Visual examination revealed no anomalies, tearing or damage.Functional testing was performed to detect any possible leakage.The anesthesia breathing circuit and breathing bag passed leak testing.The complaint was not confirmed.These devices are 100 percent in house leak tested prior to release for distribution.While no definitive root cause could be determined, this investigation revealed no intrinsic evidence to suggest a root cause related to manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® ADULT DISPOSABLE ANESTHESIA BREATHING CIRCUIT
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS HEALTHCARE MFG
s.a. de c.v. ave calidad no. 4
parque industrial internaciona
tijuana, b.c. 22425
MX   22425
Manufacturer Contact
lisa perz
6000 nathan lane north
minneapolis, MN 55442
7633833074
MDR Report Key6376799
MDR Text Key69030916
Report Number3012307300-2017-00625
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC37101307J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/27/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-