Model Number H7493952838300
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.(b)(4).
 
Event or Problem Description
It was reported that a foreign matter was found.The target lesion was located in a coronary artery.A 3.00x38mm promus premier¿ drug-eluting stent was advanced but failed to cross the lesion.After several attempts, it was noted that there was a chuck of plastic obstructing the hub.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the stent delivery system (sds) was returned for analysis.A visual examination of the stent found no issues on the stent.There were no signs of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent od (outer diameter) was measured and the result was within maximum crimped stent profile measurement.The bumper tip profile of the device showed no signs of damage.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination found no issues with the hypotube profile.The manifold hub of the device was visually examined and no issues were noted with external section of manifold hub and the threads showed no signs of damage.The internal surface of the leur section of the manifold hub appeared to be crushed with a rough edge in one section.The manifold hub was attached to a hand held encore inflation device with no issue noted with the threading.A visual and tactile examination of the outer and mid-shaft section found no issues with the extrusion profile.The bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
 
Event or Problem Description
It was reported that a foreign matter was found.The target lesion was located in a coronary artery.A 3.00x38mm promus premier¿ drug-eluting stent was advanced but failed to cross the lesion.After several attempts, it was noted that there was a chuck of plastic obstructing the hub.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER¿
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key6377190
Report Number2134265-2017-01913
Device Sequence Number16375051
Product Code NIQ
UDI-Device Identifier08714729845027
UDI-Public(01)08714729845027(17)20180508(10)19967459
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P110010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2018
Device Model NumberH7493952838300
Device Catalogue Number39528-3830
Device Lot Number19967459
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/07/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/03/2017
Supplement Report FDA Received Date04/24/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-