• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. LEAD MODEL 304

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. LEAD MODEL 304 Back to Search Results
Model Number 304-20
Device Problem High impedance (1291)
Patient Problem Seizures (2063)
Event Date 02/06/2017
Event Type  malfunction  
Event Description
It was reported that a patient had high impedance on his vns system.The patient's generator had been replaced about three months prior to the reported high impedance.Clinic notes were received, and the high impedance was seen during interrogation and diagnostics.The patient had stopped feeling stimulation.X-rays were performed, but they were not released to the manufacturer.The device history records of the generator and lead were reviewed, and both devices conformed to all specifications prior to release.No further information has been received to date, and no known surgical intervention has occurred to date.
 
Event Description
The patient had lead revision surgery due to high impedance.The impedance was tested with the existing lead reinserted into the generator, but high impedance was still present.Therefore, the lead was replaced, and the high impedance resolved.The explanting facility discarded the lead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 304
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6377216
MDR Text Key69045853
Report Number1644487-2017-03311
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2017
Device Model Number304-20
Device Lot Number3788
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 02/06/2017
Initial Date FDA Received03/03/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/12/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age37 YR
-
-