It was reported that a patient had high impedance on his vns system.The patient's generator had been replaced about three months prior to the reported high impedance.Clinic notes were received, and the high impedance was seen during interrogation and diagnostics.The patient had stopped feeling stimulation.X-rays were performed, but they were not released to the manufacturer.The device history records of the generator and lead were reviewed, and both devices conformed to all specifications prior to release.No further information has been received to date, and no known surgical intervention has occurred to date.
|
The patient had lead revision surgery due to high impedance.The impedance was tested with the existing lead reinserted into the generator, but high impedance was still present.Therefore, the lead was replaced, and the high impedance resolved.The explanting facility discarded the lead.
|