| Model Number |
ACT1530 |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/01/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Covidien reference #: (b)(4).Date of initial report: 03/03/2017.This complaint is part of a retrospective review as part of a remediation related to capa(b)(4).Medtronic¿s investigation identified the root cause to be inadequate solder joint between the device¿s thermocouple and inner tube.All identified corrective actions have been implemented including retraining of the operators on the soldering process, revalidation of the soldering process, the addition of a new electrical resistance tester, and revalidation of the thermocouple test process.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device could not sense the correct temperature and the temperature was fluctuating on the monitor.The issue was found prior to use and there was no patient injury.
|
| |
|
Search Alerts/Recalls
|