Model Number 9
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/04/2014
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event unit was returned for evaluation.Upon observation, engineering noticed the inner ring was flipped in the wrong direction.The root cause of the customer's experience was the wrong direction of the flips.It is likely a team member manually flipped the inner ring in the wrong direction.Production has been retrained.Applied medical will continue to monitor its vigilance system for trends and take appropriate actions as necessary.This report is being filed as a result of a re-review of applied medical complaints received between june 1, 2014 and may 31, 2016.This retrospective review was associated with a quality management system (qms) compliance action plan developed and presented to fda to address an april 10, 2015 warning letter.Applied medical has revised its mdr reporting criteria to be more conservative and has improved complaint handling and mdr reporting processes.The reviews ensured that recent reportable events were appropriately identified and reported to the designated regulatory authority(ies).This report, which represents the initial and final reports combined, is being submitted based on the findings of that retrospective review.In accordance to 21 cfr 803.56, if additional information is obtained which was not known or was not available when this report was submitted, then a supplemental report will be submitted to the fda.
 
Event or Problem Description
Total abdominal hysterectomy - "alexis flexible ring wouldn't conform to patient anatomy and kept popping out." patient status - unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G6313, L ALEXIS O C-SEC RET 5/BX
Common Device Name
KGW
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6378860
Report Number2027111-2017-01382
Device Sequence Number18951072
Product Code KGW
UDI-Device Identifier00607915118129
UDI-Public(01)00607915118129(17)170313(30)01(10)1217743
Combination Product (Y/N)N
Initial Reporter CountryZA
PMA/510(K) Number
9
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/13/2017
Device Model Number9
Device Catalogue Number9
Device Lot Number1217743
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2015
Distributor Facility Aware Date02/03/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2015
Initial Report FDA Received Date03/04/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-