| Model Number |
A3213 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/06/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Ra has just received the incident device and has been assigned to engineering for evaluation.A follow-up report will be sent upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed unknown - "(on (b)(6) 2017: during procedure) ratchets of the clamp a3213 were disengaged during a procedure.The ratchets pawls didn't engage to the other side properly.Medical staff confirmed that the ratchet pawls didn't match each other properly while preparing the procedure.This didn't affect the patient." type of intervention: unk.Patient status: no health damage was reported for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: the event unit was returned to applied medical for evaluation.Testing confirmed the locking mechanism was functional.Upon further investigation, it was observed that the first four ratchets of the event unit were loose and were able to separate with minimal force.It is likely that ratchet teeth were worn due to normal use over time, which would cause reduced engagement of the locking mechanism.It is likely the event unit had been used multiple times prior to the complainant's experience.Applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.
|
| |
|