The device, intended for use in treatment, was returned for evaluation.There was no relationship found between the returned device and the reported incident.Visual inspection revealed a few minor scratches on the exterior of the returned device.Functional evaluation revealed the subject controller performed as intended and exhibited no functionality issues during the testing process.At no time during testing did the subject controller show any kind of error message or fail to activate a known good wand several times without incident.All of the ablation and coag voltage output readings were within specified ranges when the subject controller was tested.The complaint could not be verified and a root cause could not be determined in association with the subject controller due to the subject unit functioning as intended when tested.Precautionary statements in the user's manual include: a thorough understanding of the principles and techniques involved in electrosurgical procedures is essential to avoid shock and burn hazards to both patient and medical personnel and damage to the device and other medical instruments.Ensure that insulation or controller grounding is not compromised.Do not wrap patient cable around metal objects.Wrapping cords around metal objects may induce currents that could lead to shocks, fires, or injury to the patient or surgical personnel.Do not allow patient contact with grounded conductive objects, such as a surgical table frame or an instrument table, to avoid risk of shock.Grounding pads should not be used.Do not contact metal objects with an activated wand.As with other electrosurgical units, electrodes and cables can provide paths for high frequency current.Position the cables to avoid contact with the patient or other electrical leads.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|