ANGIODYNAMICS VORTEX CT PORT ACCESS SYSTEM; SMARTPORT CT Back to Search Results
Catalog Number H787CT96STSD0
Medical Device Problem Codes Difficult to Flush (1251); Occlusion Within Device (1423); Difficult to Remove (1528); Device Issue (2379)
Health Effect - Clinical Codes Infiltration into Tissue (1931); No Known Impact Or Consequence To Patient (2692)
Date of Event 02/07/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that the device involved in the incident is available to be returned to the manufacturer for evaluation.To date the device has yet to be returned.Attempts are being made to obtain the device.An investigation into the root cause for incident is currently in progress.The results of the investigation will be sent via a follow up medwatch.Complaint reference (b)(4).Device not returned to date.
 
Event or Problem Description
As reported to angiodynamics on (b)(6) 2017: port failure due to fibrin sheath.Port had to be removed because it wouldn't aspirate or infuse.It was reported defective device is available for return to the manufaturer for a device evaluation.
 
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description is not confirmed, as no sample was returned for evaluation.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.A device history review of the packaging, port assembly and catheter tubing lots revealed no quality related issues or manufacturing deficiencies at the time of manufacture.The instructions for use, which is supplied to the user with this catalog number, contains the following statements: - absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.- if the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Potential complications: catheter fragmentation and catheter pinch-off.Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Pinch-off syndrome: pinch-off syndrome signs may include difficulty in aspirating blood, resistance to flushing or infusion of medications or fluids that improves with position changes, infraclavicular pain and/or swelling with catheter flushing or infusion palpitations, sudden onset chest pain, cardiac arrhythmias, extra heart sound, chest wall swelling at the port pocket, vein insertion site, pain in shoulder or port area not associated with swelling, cough, paresthesia of arm on side of catheter withdrawal occlusion or swishing sound with catheter flushing.Warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Note: if infusion or aspiration is successful upon lifting arm above the head and turning the head, consider pinch-off syndrome as a possible cause.The line should be radiologically evaluated if pinch-off syndrome is suspected.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VORTEX CT PORT ACCESS SYSTEM
Common Device Name
SMARTPORT CT
Manufacturer (Section D)
ANGIODYNAMICS
one horizon way
manchester GA 31816
MDR Report Key6381118
Report Number1056436-2017-00024
Device Sequence Number442387
Product Code LJT
UDI-Device IdentifierH787CT96STSD0
UDI-PublicH787CT96STSD0
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K062414
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 08/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Catalogue NumberH787CT96STSD0
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2017
Supplement Date Received by Manufacturer02/07/2017
Initial Report FDA Received Date03/06/2017
Supplement Report FDA Received Date08/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-