CODMAN & SHURTLEFF, INC. / MEDOS S.A. PROGRAMMABLE VALVE INLINE; SHUNT, CENTRAL NERVOUS SYSTEM & COMPS Back to Search Results
Catalog Number 82-3834
Device Problem Device Operational Issue (2914)
Patient Problem No Information (3190)
Event Date 02/13/2017
Event Type  Injury  
Manufacturer Narrative
Udi: (b)(4).Upon completion of the investigation a follow up report will be filed.
 
Event Description
As reported by sales rep, a chpv appeared to be malfuctioning.Valve was replaced.Patient's condition after event described as stable; no delays over 30 minutes reported.Rep is returning valve and requests an investigation report for the surgeon.
 
Manufacturer Narrative
(b)(4).Upon completion of the investigation it was noted that the images were taken of the valve ¿as received¿.The valve was visually inspected: no defects were noted.The position of the cam when valve was received was 30mmh2o.The valve was hydrated.The valve was tested for programming with programmer 82-3126 with serial number (b)(4), the valve passed the test.The valve was flushed, the valve passed the test no occlusion was noted.The valve was leak tested, no leaks were noted.The valve was reflux tested.The valve passed the test.The valve was dried.The valve was then pressure tested, the valve passed the test.Review of the history device records confirmed the valve product code 82-3834, with lot crfb3j, conformed to the specifications when released to stock on the 23rd may 2014.No root cause could be determined, as the problem reported by the customer could not be duplicated.Based on the results of this investigation no further action is required.Trends will be monitored for this and similar complaints.At the present time this complaint is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMABLE VALVE INLINE
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM & COMPS
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC. / MEDOS S.A.
rue girardet 29
le locle CH 24 00
SZ  CH 2400
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key6381140
MDR Text Key69171330
Report Number1226348-2017-10142
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K974739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2019
Device Catalogue Number82-3834
Device Lot NumberCRFB3J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/23/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age60 YR
-
-