• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CB010, 5MMX34CM SCISSOR REPOS CRTR 10/BX; HET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CB010, 5MMX34CM SCISSOR REPOS CRTR 10/BX; HET Back to Search Results
Model Number 6
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/20/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event unit was returned for evaluation.Upon inspection, engineering was able to confirm the customer experience.The root cause of poor blade performance is a loose screw.Engineering was unable to determine when or why the screw loosened.Production was retrained on the proper screw adjustment as a precautionary measure.Applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.This report is being filed as a result of a re-review of applied medical complaints received between june 1, 2014 and may 31, 2016.This retrospective review was associated with a quality management system (qms) compliance action plan developed and presented to fda to address an april 10, 2015 warning letter.Applied medical has revised its mdr reporting criteria to be more conservative and has improved complaint handling and mdr reporting processes.The reviews ensured that recent reportable events were appropriately identified and reported to the designated regulatory authority(ies).This report, which represents the initial and final reports combined, is being submitted based on the findings of that retrospective review.In accordance to 21 cfr 803.56, if additional information is obtained which was not known or was not available when this report was submitted, then a supplemental report will be submitted to the fda.
 
Event or Problem Description
Laparoscopy - incident 2 - "when attempting to cut tissue, the scissor would not cut through.This is the 2nd unit from same batch, same box.Returned all remaining lot no.Of product.Patient status - unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CB010, 5MMX34CM SCISSOR REPOS CRTR 10/BX
Common Device Name
HET
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6382348
Report Number2027111-2017-01397
Device Sequence Number13591411
Product Code HET
UDI-Device Identifier00607915117320
UDI-Public(01)00607915117320(17)170901(30)01(10)1230275
Combination Product (Y/N)N
Initial Reporter CountryAU
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/01/2017
Device Model Number6
Device Catalogue Number6
Device Lot Number1230275
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2015
Initial Report FDA Received Date03/06/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-