ABBOTT MEDICAL OPTICS TECNIS CL; MONOFOCAL IOLS Back to Search Results
Model Number Z9002
Medical Device Problem Codes Difficult To Position (1467); Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/08/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to the manufacturer has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was removed and replaced during the initial procedure.This lens was a replacement lens for an earlier lens explant.During the procedure, the lens went into the eye upside down.The doctor tried to fix it but the lens could not seem to stay centered in the bag.The lens continued to be decentered.This lens was removed and replaced with a non-amo device.There was no incision enlargement and no vitrectomy or patient post-op injury reported.No further information was provided.This report captures serial number (b)(4) and a separate mdr report will be filed for the explanted lens.
 
Additional Manufacturer Narrative
Device available for evaluation ¿ yes, returned to manufacturer on 04/10/2017 device returned to manufacturer ¿ yes.Device evaluation: the device was returned to the manufacturer.Device inspection was performed and the lens was observed under microscope where viscoelastic residue and particles were observed.No damage was observed in the lens optic zone or in the haptics.The customer's reported issue could not be confirmed in the returned lens sample.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no other complaints were received for this order number to date.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the device.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6382641
Report Number2648035-2017-00419
Device Sequence Number528483
Product Code HQL
UDI-Device Identifier05050474530645
UDI-Public(01)05050474530645(17)210222
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date02/22/2021
Device Model NumberZ9002
Device Catalogue NumberZ900200175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/08/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/06/2017
Supplement Report FDA Received Date05/30/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
-
-