MAUDE Adverse Event Report: STRYKER ORTHOBIOLOGICS-MALVERN VITOSS BIOACTIVE FOAM PACK 5CC; RESORBABLE CALCIUM SALT BONE VOID FILLER DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOBIOLOGICS-MALVERN VITOSS BIOACTIVE FOAM PACK 5CC; RESORBABLE CALCIUM SALT BONE VOID FILLER DEVICE Back to Search Results
Catalog Number 2102-1605
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/08/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that; during a follow up visit the patient presented with an infection.Product was protruding out of the wound.
 
Additional Manufacturer Narrative
Method: device history review, complaint history review, risk assessment; result: device history review indicated all devices released into final stock met specifications.Device evaluation could not be performed as no items were returned.Conclusion: the root cause could not be determined conclusively and is likely multifactorial.
 
Event or Problem Description
It was reported that; during a follow up visit, the patient presented with an infection.Product was protruding out of the wound.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITOSS BIOACTIVE FOAM PACK 5CC
Common Device Name
RESORBABLE CALCIUM SALT BONE VOID FILLER DEVICE
Manufacturer (Section D)
STRYKER ORTHOBIOLOGICS-MALVERN
45 great valley parkway
malvern PA 19355
MDR Report Key6384294
Report Number0002530131-2017-00003
Device Sequence Number317000
Product Code MQV
UDI-Device Identifier00808232000511
UDI-Public(01)00808232000511
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date04/28/2018
Device Catalogue Number2102-1605
Device Lot NumberB1604004
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/08/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/07/2017
Supplement Report FDA Received Date06/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age59 YR
-
-