| Catalog Number |
H787107323015 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260); Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/13/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported defective device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
|
| |
|
Event or Problem Description
|
|
As reported to angiodynamics on (b)(6) 2017: when opening the sterile packaging during prep for an angiographic procedure it was noted the tip of the angiographic catheter had fractured off inside of the packaging.The device was set aside and a new of the same device was used to complete the procedure.The reported defective disposable device has been returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned for evaluation was one soft-vu catheter.A visual review of the catheter noted that the tip was fractured,.3cm from the weld joint.The customer's reported complaint description of soft-vu tip fracture is confirmed.The defect was discovered prior to making patient contact.Although the complaint description is confirmed, a root cause for the defect cannot be determined.A review of the lot history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, which is supplied to the end user with this catalog number, contains a statement; "reshaping of catheter tip is not recommended.Physical damage to the catheter material can result when exposed to heat".Capa (b)(4) was initiated to determine the root cause and implement a corrective / preventative action for this type failure.A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
|
| |
|
Search Alerts/Recalls
|