| Model Number |
METRIX AIR |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/13/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause of malfunction: user had an inaccurate reference.(b)(4).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.The customer is concerned with tests results from results obtained of 89 mg/dl.The customer's expected fasting blood glucose test result range is 150 - 160 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2017, a back to back blood test was not performed by the customer; the customer declined.The product is stored according to specification.The test strip lot manufacturer's expiration date is 04/30/2018 and open vial date at time of call is a few days.The meter memory was reviewed for previous test result history (date/time not set): (b)(4).
|
| |
|