Catalog Number 42955
Medical Device Problem Code Occlusion Within Device (1423)
Health Effect - Clinical Code Therapeutic Response, Decreased (2271)
Date of Event 02/04/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The returned valve was patent.The valve met the requirements for siphon, reflux, and pre-implantation testing.The valve did not meet the requirements for pressure-flow testing.There was proteinaceous debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open resulting in fluid reflux and affecting the flow of fluid through the valve.Instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ the returned valve also did not meet the requirements for leak due to a tear in the top of the reservoir.It is unknown how or when this damage occurred.The ifu caution that ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting, or crushing of components.¿ the ifu also caution that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance.A review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.
 
Event or Problem Description
It was reported to medtronic neurosurgery that the device stopped draining.According to the report, the device was explanted and there was no injury.
 
Additional Manufacturer Narrative
Additional information received reported the patient had undergone an mri and adjustment.In addition, the valve had been soaked in bacitracin prior to implant.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAMR VALVE, SMALL
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key6390023
Report Number2021898-2017-00090
Device Sequence Number10900298
Product Code JXG
UDI-Device Identifier00643169692718
UDI-Public00643169692718
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K152700
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Catalogue Number42955
Device Lot NumberE12556
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2017
Initial Date Received by Manufacturer 02/06/2017
Supplement Date Received by ManufacturerNot provided
03/31/2017
Initial Report FDA Received Date03/08/2017
Supplement Report FDA Received Date04/28/2017
10/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age32 MO
Patient Weight11
-
-