It was reported that at 1500 on (b)(6) 2017, the patient was receiving a lower limb infusion.The bd q-syte¿ luer access split-septum stand-alone device was being used with an unknown non-bd brand sti and infusion set.The nurse found the suspect device broken at 1830 on (b)(6) 2017 and noted blood and medication had leaked from the connector.The medication being infused was methylprednisolone, mucosolvan, and avelox dissolved in 5% glucose saline.The middle part of the suspect device was found broken in two section.A new q-syte was immediately replaced and the infusion was continued.At this time, the patient's blood results were normal, vital signs steady, and the doctor estimated the blood loss to be about 30 ml.The patient died on (b)(6) 2017.The patient was initially hospitalized for copd but was not critically ill.Per report on (b)(6) 2017, "at present, the hospital did not identify the final death of patients but the hospital may be considered as a cause of cardiac disease." no additional information is available.
|