Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information of conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
|
According to the available information, for the surgery a pentax endoscope and a dormia were used.They captured the stone, but it was not able to pass the papilla.Due to this they continued with the stone fragment device (the stone fragment device is a mechanical lithotriptor).During compression, the wire of the basket broke.Nevertheless, the fragmented stone could be removed with the broken basket and the endoscope.This resulted in a longitudinal crack of the duodenal wall with strong bleeding.Hemostasis was not possible.Due to this 2nd surgery (laparotomy) at the same day.After 2 more re-laparotomies and closure of fistula the patient improved.He was released from the hospital to outpatient treatment - subjectively he felt good.It is not clear which lot of the dormia was used, lots 10050286 and 10070002.Both lots expired in 2015.Dor424 is not indicated for biliary extraction, and for use with any lithotriptor.Device was probably expired.
|