| Model Number |
0832F-00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
02/07/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The customer discarded the kit and did not return it for the investigation.A lot history review was completed.This lot was manufactured according to approved procedures and met all specifications for release, with no noted manufacturing deviations or non-conforming events.The root cause of the reported complaint can't be determined.There is no sample or photo available to determine if there was a defect related to the manufacturing of the affected item or if user error was the cause of the issue.
|
| |
|
Event or Problem Description
|
|
Haemonetics was notified on 02/09/2017 of a needlestick event that occurred with the collection set, (b)(4), filtered, utilized with a collection on the mcs+ 8150 apheresis system.The user was processing the collection set post donation and engaged the needle guard by pulling it.The user thought she heard the needle guard click and lock, but it had not locked and the user was stuck with the used needle.The user was seen at a local clinic and was given a tetanus shot as a precautionary measure.The user stated her last tetanus shot was seven years prior to this event.The user went back to the clinic for a follow up appointment one week later with no health issues.
|
| |
|