On march 18, 2016, applied medical issued a voluntary class ii recall of the ca090 direct drive clip applier due to increased customer feedback indicating inconsistent clip application, which has the potential to lead to unoccluded vessels.An internal corrective and preventative action (capa) has been initiated to conduct a thorough investigation and implement appropriate corrective actions.Fda has issued recall number z-1621-2016 for this recall.This report is being filed as a result of a re-review of applied medical complaints received between june 1, 2014 and may 31, 2016.This retrospective review was associated with a quality management system (qms) compliance action plan developed and presented to fda to address an april 10, 2015 warning letter.Applied medical has revised its mdr reporting criteria to be more conservative and has improved complaint handling and mdr reporting processes.The reviews ensured that recent reportable events were appropriately identified and reported to the designated regulatory authority(ies).This report, which represents the initial and final reports combined, is being submitted based on the findings of that retrospective review.
|
|
Lap chole - "the staff opened a clip applier and attempted to use it in the patient, it only had three clips in it which were attempted to be used in the patient.Upon query to the staff nurse michael informs that they needed to open another device in order to finish the procedure." patient status - unknown, patient status unknown.
|