| Catalog Number |
03.010.228 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/16/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient weight is not available for reporting.Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.A device history record (dhr) review was conducted for part # 03.010.228, lot # us99067: release to warehouse date: 24-nov-2008, expiration date: na, supplier: (b)(4).No non conformance reports (ncrs) were generated during production.Review of the device history records showed that there were no issues during the manufacture of this product that would contribute to this complaint condition.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure for a right femur fracture on (b)(6) 2017, two drill bits broke.As the surgeon was drilling a path for the recon screws with a synthes small battery driver, he approached the medial calcar and an inch of the drill bit broke off in the femoral neck region.The surgeon tried again with another drill bit and an inch of the second drill bit broke off in the femoral neck region.Approximately an inch of both drill bit tips were left in the patient.The surgeon did not attempt to retrieve the tips of the drill bits.It would have caused more damage to the patient to retrieve the tips.There was a 40 minute surgical delay.Routine intraoperative x-rays were taken.The surgeon could not place the inferior screw and had to abandon the intended procedure.The surgeon had already placed the adolescent femoral nail and the superior screw (5.0 mm recon screw), but was unable to place the inferior screw after the tips of the drill bit broke.The surgeon completed the procedure having preferred to have implanted the second (inferior screw).The patient was reported as stable.Concomitant device reported: synthes small battery driver (quantity of 1).This is report 2 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.A product development investigation was performed for the subject device (drill bit for 5.0mm recon screws/large qc, part number 03.010.228, lot number us99067).The subject device was returned with the complaint condition stating: two 03.010.228 lot number us99067 drill bits for 5.0mm recon screws were returned and reported to have broken during surgery with fragments retained in the patient.This complaint condition was likely caused by over eight years of consistent use and possible rough handling during surgery or sterile processing; however, this complaint is not likely a result of any design or manufacturing related deficiency.A visual inspection, functional test, complaint history review, drawing review, and risk assessment review were performed as part of this investigation.This complaint is confirmed.The 03.010.228 drill bit for 5.0mm recon screws is an instrument routinely used in the titanium cannulated adolescent lateral entry femoral nail system.The devices were returned and reported to have broken during surgery with fragments retained in the patient.This condition is confirmed; the distal step of both drill bits is missing and x-ray evidence confirms their placement in the patient¿s femur.It is likely that over eight years of consistent use and possible rough handling during surgery or sterile processing has led to this complaint condition.The devices were manufactured in 11/2008 and are over eight years old.The balance of the returned devices is in fairly worn condition consistent with the device¿s age.Drawing was reviewed and determined to be suitable for the intended design, application and dimensional conformity when used as recommended.The condition of the returned device does agree with the complaint description.Whether the complaint condition for this device can be replicated is not applicable for this condition.The risk assessment does not adequately address the complaint event.Upon review of the device history record, no ncrs germane to the complaint condition were generated during the production of this device.It is likely that over eight years of consistent use and possible rough handling during surgery or sterile processing has led to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|