| Catalog Number |
1012452-12 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/11/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Concomitant medical products: guide wire: run through copilot bleedback control valve guide catheter: 6 fr xp.The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure of the heavily calcified, moderately tortuous lesion in the mid left anterior descending coronary artery, the nc trek was advanced, but would not inflate.The device was removed and replaced with another nc trek to successfully complete the procedure.After the case, inflation was tested outside the anatomy, but it still would not inflate.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided regarding this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal file number - (b)(4).Evaluation summary: the device returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda on march 17, 2017 [medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.The abbott internal recall number is 2024168-3/14/2017-002-r.
|
| |
|
Search Alerts/Recalls
|