|
The returned valve was patent.It met the requirements for reflux, pressure-flow, pre-implantation.Proteinaceous debris was noted within the interior and exterior of the valve.The instructions for use that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization or other debris¿.The valve did not meet the requirements for leak testing due to a tear in the top of the cassette housing chamber.It is unknown how or when this damage occurred.The instructions for use that accompany the device caution that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance.¿ it is also suggested that the valve be placed in a surgically created loose subgaleal pocket.This allows gentle placement of the valve and minimizes handling with surgical tools.A review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.
|