ABBOTT MEDICAL OPTICS TECNIS 1-PIECE; MONOFOCAL IOLS Back to Search Results
Model Number ZCB00
Medical Device Problem Code Sticking (1597)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No indication of lens explant.(b)(6).Device evaluation: the lens was not returned as it remains implanted.Therefore, the customer's reported issue could not be verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provide instructions and precautions along with warnings for the proper use and handling of the device.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that the haptic stuck to the optic and the haptics stuck to each other.The intraocular lens (iol) was implanted.There was no patient injury.No additional information was provided to abbott medical optics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6397624
Report Number2648035-2017-00467
Device Sequence Number13460414
Product Code HQL
UDI-Device Identifier05050474531864
UDI-Public(01)05050474531864(17)201028
Combination Product (Y/N)N
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/28/2020
Device Model NumberZCB00
Device Catalogue NumberZCB0000275
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2017
Initial Report FDA Received Date03/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-