| Catalog Number |
9734302 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/13/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient was involved in this concern.Part was not replaced or returned despite multiple attempts by the manufacturer to complete the return materials authorization process.This device is included in the medical device field correction notification, "potential for user injury ¿ medtronic navigated cannulated taps" (may 2015).The revised instructions for use (ifu) were also provided with the notification.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that a solera 6.5 tap had a guide wire stuck inside.No patient or user harm reported.This issue was discovered outside of surgical use.
|
| |
|