• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER ACCESS FREE T4 REAGENT; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER ACCESS FREE T4 REAGENT; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 33880
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/22/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The customer did not supply the patients' demographics such as date of birth and weight.A beckman coulter (bec) field service engineer (fse) was not dispatched to the customer's site.No hardware errors, flags or other assay issues were reported in conjunction with this event.There is no evidence that the access free t4 reagent was returned for evaluation.The cause of this event cannot be determined with the available information.(b)(4).All mdrs associated with this report are: 2122870-2017-00022, 2122870-2017-00023.
 
Event or Problem Description
The customer contacted beckman coulter on march 06, 2017 to report generating reproducible, elevated free thyroxine (access free t4) and free triiodothyronine (access free t3) results for one patient on the unicel dxi 800 access immunoassay system serial numbers (b)(4).The customer could not provide information on which instrument was in use at the time of this event.The elevated results did not correlate with the reproducible thyroid-stimulating hormone (access tsh) results generated from the same patient samples, which recovered with one result below the assay's normal reference range and all other samples recovering within the access tsh normal reference range.The elevated access free t4 and access free t3 results also did not correlate with results obtained on an alternate methodology (diasorin liaison).All seven samples from the patient generated either an elevated access free t4 result, an elevated access free t3 result or an elevated result for both analytes.One accesstsh result for the patient, obtained on (b)(6) 2017, recovered below the assay's normal reference range.All other accesstsh results obtained for the patient recovered within the assay's normal reference range.A sample from the patient was sent to an alternate facility for testing on an alternate methodology (diasorin liaison xl) and lower free t4 and free t3 results, within each assay's respective normal reference range, were obtained.There was change or impact to patient care or treatment which occurred due to the reproducible elevated access free t3 and access free t4 patient results, as the patient started thyreostatic drug therapy.Quality control (qc) and calibrations were performing within assay and instrument specifications at the time of the event.Patient sample collection and processing information such as tube type and centrifugation, was not provided by the customer.This report addresses the elevated access free t4 results obtained for this patient.Mdr 2122870-2017-00022 addresses the elevated access free t3 results obtained for this patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS FREE T4 REAGENT
Common Device Name
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
BECKMAN COULTER
1000 lake hazeltine drive
chaska MN 55318
MDR Report Key6402373
Report Number2122870-2017-00023
Device Sequence Number967965
Product Code CEC
UDI-Device Identifier15099590225834
UDI-Public(01)15099590225834(17)180731(11)160817(10)671100
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
982250
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2017
Device Catalogue Number33880
Device Lot Number671100
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2017
Initial Report FDA Received Date03/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age14 MO
-
-