| Catalog Number |
4004C0513 |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Material Protrusion/Extrusion (2979); Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/09/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The device was implanted into the patient.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure using penumbra coil 400''s (pc400's).After deploying and detaching the last pc400 of the procedure, the physician removed the px slim delivery microcatheter (px slim) out of the patient and subsequently, the end of the pc400 migrated down into the parent artery.The physician then attempted to use a stent device but was unable to successfully place it.Therefore, the physician left the single loop of the pc400 protruding out of the aneurysm since he did not believe that it would cause any problems.The patient was then placed on plavix and the procedure was completed.There was no report of an adverse effect to the patient.
|
| |
|