| Model Number |
60-02-15 |
| Medical Device Problem Codes |
Device Operational Issue (2914); Physical Property Issue (3008)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).This medwatch report is being filed on behalf of (b)(4).A (b)(4) field service representative was dispatched to the facility to investigate.The service representative tightened the guide sleeve (60-02-02) of the encoder to resolve the reported issue.Subsequent testing found no further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Evaluated on site by (b)(4) technician.
|
| |
|
Event or Problem Description
|
|
(b)(4) received a report that the encoder of the centrifugal pump system with tubing clamp would not control the flow speed during priming.The user stated that the knob was found to be loose.There was no patient involvement.
|
| |
|