• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-FRANCE XIA II VITALIUM ROD 6.0 MM X 600 MM; PEDICLE SCREW SPINAL SYSTEM.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-FRANCE XIA II VITALIUM ROD 6.0 MM X 600 MM; PEDICLE SCREW SPINAL SYSTEM. Back to Search Results
Catalog Number 03822601
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Injury (2348); Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 02/14/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that: cobalt chrome rod broken.Blockers loose at t10, t11, l4, l5, s1, iliac bolt and offset connector.Screw loose at left s1.Multiple levels of nonunion.Replaced rod and affected screws.
 
Additional Manufacturer Narrative
Risk assessment; no product was returned and lot number was not provided, therefore device inspection, device history review and complaint history review could not be performed.Without the device inspection and further information the root cause cannot be determined conclusively.
 
Event or Problem Description
It was reported that: cobalt chrome rod broken.Blockers loose at t10, t11, l4, l5, s1, iliac bolt and offset connector.Screw loose at left s1.Multiple levels of nonunion.Replaced rod and affected screws.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA II VITALIUM ROD 6.0 MM X 600 MM
Common Device Name
PEDICLE SCREW SPINAL SYSTEM.
Manufacturer (Section D)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR  33610
MDR Report Key6408141
Report Number0009617544-2017-00096
Device Sequence Number16375106
Product Code NKB
UDI-Device Identifier04546540539656
UDI-Public(01)04546540539656
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number03822601
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/14/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/15/2017
Supplement Report FDA Received Date06/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient Weight127
-
-