• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY4C0201
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2017-00384, 3005168196-2017-00386.The hospital disposed of the device.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure on (b)(6) 2017 using ruby coils.During the procedure, while advancing a ruby coil through a non-penumbra microcatheter, the physician experienced resistance and the ruby coil became stuck and unable to advance.Although the physician experienced resistance while retracting it, the ruby coil was successfully removed.The procedure was then halted for reasons not related to any penumbra product.There was no report of an adverse effect to the patient.The physician then reattempted the coil embolization procedure on (b)(6) 2017 using ruby coils.During the procedure, while advancing a ruby coil through a non-penumbra microcatheter, the physician experienced resistance and the ruby coil was unable to advance out of the microcatheter; therefore,it was removed.This same issue occurred again with a new ruby coil, and then the procedure was completed using another new ruby coil and the same non-penumbra microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6408232
Report Number3005168196-2017-00385
Device Sequence Number13467323
Product Code HCG
UDI-Device Identifier00814548013114
UDI-Public00814548013114
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date03/29/2024
Device Catalogue NumberRBY4C0201
Device Lot NumberF68727
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2017
Initial Report FDA Received Date03/15/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age80 YR
Patient Weight59
-
-