| Catalog Number |
9734404 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/15/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A field representative reported that the tip of the awl tip instrument was bent, however, the part was not returned to manufacturer for analysis, therefore, no findings are possible.Replacement of the instrument resolved the issue.No further issues were reported.Part not returned.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that while in a spinal fusion procedure, they discovered a bent tip on the awl tip instrument.There was no reported delay to the procedure due to this issue.There was no impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
The hardware investigation of the returned awl tip found that the reported event was related to a physical damage issue.As reported, the tip of the awl was visibly bent.There are also impact marks at the back end of the instrument.The reported event was confirmed.This issue was documented in a medtronic navigation hardware anomaly tracking database.
|
| |
|