| Catalog Number |
48552322 |
| Medical Device Problem Codes |
Disassembly (1168); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the head broke off the tulip screw during the case.The further reported that the construct had been locked down with the rod and the blocker.The surgeon then checked on x-ray and wanted to change the direction of the tulip head.When he tried to adjust it the head broke off the screw.The tulip head was removed with forceps and the surgeon was able to unwind the thread from the bone.No unintended metal was left in the patient.The bone was good quality and there were no other obvious issues.A new screw was inserted and the case was completed successfully with no lasting consequences for the patient.There was a 30 minute delay to surgery as a result of the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection, device history review, complaint history review, risk assessment.The screw tulip was confirmed separated based on visual examination.Manufacturing records for this product were reviewed and no anomalies were found.Potential root causes may be: user applies too much torsional force.User applies too much bending/cantilever force.
|
| |
|
Event or Problem Description
|
|
It was reported that the head broke off the tulip screw during the case.The further reported that the construct had been locked down with the rod and the blocker.The surgeon then checked on x-ray and wanted to change the direction of the tulip head.When he tried to adjust it the head broke off the screw.The tulip head was removed with forceps and the surgeon was able to unwind the thread from the bone.No unintended metal was left in the patient.The bone was good quality and there were no other obvious issues.A new screw was inserted and the case was completed successfully with no lasting consequences for the patient.There was a 30 minute delay to surgery as a result of the reported event.
|
| |
|
Search Alerts/Recalls
|