• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-45
Device Problems Device Alarm System (1012); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/31/2017
Event Type  malfunction  
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The customer reported that the device has undergone various repairs at livanova (b)(4).However, due to the intermittency of the reported issue, the reported problem has not been resolved.The customer believes the issue is more prominent after a cold start and goes away after a while.The device has been requested for return to livanova (b)(4) for investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Not returned to manufacturer.
 
Event Description
Livanova (b)(4) received a report that the s5 gas blender system has had intermittent error and alarm issues over the past several months.There was no report of patient injury.
 
Manufacturer Narrative
Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The unit was returned to livanova (b)(4) for investigation and repair.Inspection of the device was able to reproduce the described error.During evaluation, the air and co2 gas controllers were found to require replacement.The parts were replaced and the unit was reassembled and tested.Functional verification testing was completed without further issues and the unit was returned to the customer.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6415631
MDR Text Key70341230
Report Number9611109-2017-00256
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-