MAUDE Adverse Event Report: SYNTHES MEZZOVICO 2.7MM/3.5MM VA-LCP EXT MEDL DSTL HUM PL 6H/LT/137MM-XLNG; PLATE,FIXATION,BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES MEZZOVICO 2.7MM/3.5MM VA-LCP EXT MEDL DSTL HUM PL 6H/LT/137MM-XLNG; PLATE,FIXATION,BONE Back to Search Results
Catalog Number 02.117.706
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Impaired Healing (2378); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(6).Additional device product code hwc.Device is not expected to be returned for manufacturer review/investigation.(b)(4).Device history records review was completed for part# 02.117.706, lot# 8013953.Manufacturing location: (b)(4), manufacturing date: aug 06, 2012.No non conformance reports were generated during production.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that a patient was implanted with four unknown synthes 3.5 cortex screws, eight unknown 2.7 mm val screws, six unknown 3.5 locking screws, a 7 hole 3.5 mm lcp posterolateral distal humerus plate, and a 6 hole extra long 2.7/3.5 mm va-lcp extended medial distal humerus plate to treat a left distal humerus fracture on an unknown date.On an unknown date postoperatively, it was identified that the patient had developed a non-union of the fracture.On (b)(6) 2017, the patient was returned to the operating room and all of the devices were removed easily and intact.The non-union was debrided and grafted and the patient was revised with longer plates.No delay in surgery was reported and no adverse events against the patient were identified.Procedure was completed successfully and patient outcome was reported as okay.This report is for one (1) 2.7 mm/3.5 mm va-lcp ext medl dstl hum pl 6h/lt/137 mm-xlng.This is report 2 of 5 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM/3.5MM VA-LCP EXT MEDL DSTL HUM PL 6H/LT/137MM-XLNG
Common Device Name
PLATE,FIXATION,BONE
Manufacturer (Section D)
SYNTHES MEZZOVICO
zona industriale 4
mezzovico CH680 5
SZ  CH6805
MDR Report Key6416944
Report Number1000562954-2017-10030
Device Sequence Number38101
Product Code HRS
UDI-Device Identifier10886982037268
UDI-Public(01)10886982037268(10)8013953
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K120070
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Catalogue Number02.117.706
Device Lot Number8013953
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2017
Initial Report FDA Received Date03/20/2017
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight99
-
-