| Model Number |
ENBF2516C166E |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Rupture (2208)
|
| Date of Event |
02/25/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant products: enlw1620c93e sn (b)(4), expiration date: 2013-12-21, udi # (b)(4).Enew2020c82e sn (b)(4), expiration date: 2014-01-05, udi # (b)(4).A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant stent graft system was implanted in a patient for the endovascular treatment of an abdominal aortic aneurysm.It was reported that the patient presented emergently with abdominal pain.A non-contrast ct was done and the patient was diagnosed with a ruptured abdominal aortic aneurysm.The patient was then taken to the operating room.An angiogram in the or showed only a type ii endoleak.The physician decided to implant an aui device and perform a femoral-femoral bypass.A 24x14x102 device was implanted in the existing stent graft on the right ipsilateral side.It was noted that it was landed 1cm proximal to the previous stent graft.A 16x16x93 extension was also implanted along with a talent occluder in the contralateral limb on the left side.An angiogram then revealed a proximal type i endoleak, so the physician decided to use another manufacturer¿s stent.The physician placed the first stent low so they used a second one.Per the physician, the cause of the ruptured aneurysm was unknown and the cause of the proximal type i endoleak, after the aui device was implanted, was also unknown.No additional clinical sequelae were reported and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Correction: a 28x14x102 device was implanted in the existing stent graft on the right ipsilateral side.It was noted that it was landed 1cm proximal to the previous stent graft, which was ~ 8mm below the lowest left renal artery.The event was reportedly resolved and the procedure was completed.The next day, the patient's creatinine had increased from 2.3 pre-procedure to 3.The patient was not making any urine.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information: sex, date of birth.If information is provided in the future, a supplemental report will be issued.
|
| |
|