| Catalog Number |
ASC860SV |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
02/21/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device was returned to manufacturing facility for analysis.Slight kinks were evident along the catheter of the 8.0mmx60mm delivery system.The balloon returned deflated with crimp impressions visible on the exposed balloon surface.There was no deformation to the distal tip.The still cine images capture the successful deployment of an assurant cobalt 8x30mm and assurant cobalt 8x60mm stent.Foreign material can clearly be seen distal to the assurant cobalt 8x60mm stent.A third assurant cobalt 8x30mm stent is used to jail the foreign material to the vessel wall.
|
| |
|
Event or Problem Description
|
|
The physician intended to use an assurant cobalt bare metal stent to treat a lesion located in the rt common/external iliac.The lesion reported to be severely calcified, moderately tortuous with 75% stenosis.Calcified stenosis of rt external artery was also noted.Prior to the procedure, the device was inspected with no issues noted.The device was prepped per the ifu and no issues were identified.It was reported that during the procedure, the physician successfully implanted one assurant cobalt (8x30mm ) in the rt ostial iliac, and then when attempting to implant a second assurant cobalt (8x60mm), the physician found a foreign body distal to the stent in the vessel.The physician deployed the second stent, asked for a snare but was never used and then used a third assurance cobalt (8x30mm) stent to jail the foreign body against the arterial wall and ended the case.The physician then requested a mynx closure to close the puncture site.No further clinical sequelae was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
No foreign material was present on the returned delivery systems.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|