The device was tested following our quality control procedure.A visual control shows dry blood deposits visible inside and on the valve and the reservoir.Several marks can be seen on the connectors, probably due to the use of forceps during explantation.The functional control shows that air and alcohol cannot flow through the valve and that the valve cannot be adjusted.The pressure control turned out to be impossible to perform due to the state of return of the valve.Even after cleaning, the presence of permanent deposits prevents any pressure measurements on the valve.The state of return of the device makes it very difficult to establish the origin of the incident.The recommendation given in the ifu to return the product in sterile water was not followed therefore it was very dry and it was difficult to know whether the observed non-conformities were due to the device or its dryness.Nevertheless, considering the situation faced by the user and the particles observed inside the valve, around the inlet connector and the rotor, we can assume that the functioning difficulties were due to an obstruction of the valve, either due to the obstruction of the inlet connector by the ball, or to the impossibility to adjust the valve, due the deposits behind the rotor.Obstruction is a well-known side-effect of a shunting system, described in the ifu of the device and widely discussed in the scientific literature.
|