|
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The investigation is currently under way.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Visual/microscopic inspection: the sample appears to be clean.The balloon size for this product is printed on the balloon hub of the catheter and identified the return sample as a 22 mm x 4.5 cm.No holes or catheter damage was identified.No anomalies were noted to the device.Functional/performance evaluation: the patency of the guidewire lumen was tested using an in-house 0.035" guidewire and it passed without issue.The balloon was inflated to nominal pressure and rbp without issues.The balloon was then fully deflated.During deflation it was noted that the device was aspirating air.The deflated balloon was inserted into water, and a vacuum was applied to the catheter with an inflation device.Approximately 2-3 cc's of water was able to be pulled into the balloon.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: the device was returned.The balloon was inflated and deflated without issue.Although the sample was able to fully inflate and deflate without issue, the investigation is confirmed for the reported aspiration issues, as air could be heard aspirating during deflation of the device.The definitive root cause could not be determined based upon the available information.Labeling review: the current instructions for use (ifu) states: warnings and precautions: never use force when advancing or retracting intravascular device without first angiographically determining cause for any resistance.Using force may damage catheter or cause injury to the patient (such as vessel perforation).Do not exceed maximum inflation pressure indicated on label.Excess inflation pressure can cause balloon rupture and the inability to withdraw catheter through introducer sheath.Withdrawing balloon through introducer sheath may damage balloon.Balloon deflation and catheter removal: deflate balloon by drawing vacuum on the = 50 cc inflation device.Use fluoroscopy to visually confirm that balloon has fully deflated prior to withdrawing catheter.While maintaining negative pressure and guidewire position, withdraw deflated device over the guidewire and out through introducer sheath.Use of a gentle clockwise twisting motion may be used to help withdraw balloon through introducer sheath.If unusual resistance is met when attempting to withdraw balloon, position balloon in anatomical position in which it can be inflated safely.Partially inflate balloon and then deflate it.Balloon re-folding can be observed under fluoroscopy.Use of recommended contrast concentration will facilitate fluoroscopic visualization of balloon re-folding.If balloon will not pass through introducer sheath for removal, remove balloon and sheath as a single unit.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|