Model Number 3389S-40
Medical Device Problem Codes Break (1069); Device Damaged Prior to Receipt by User (1570)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/22/2017
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a health care provider (hcp) via a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator for parkinson's dual.It was reported that the lead was opened and discovered to be bending prior to use.It was then noted that the lead was implanted on (b)(6) 2017 and explanted on (b)(6) 2015.Follow-up is being conducted to obtain clarification on the discrepant information.There was no known impact or consequence to the patient.No further complications were reported/anticipated.
 
Additional Manufacturer Narrative
The lead was bent near the handle approximately 37 cm from the distal end.
 
Additional Manufacturer Narrative
Analysis of the lead (lot # va1cdnk) found that the #0 electrode was bent.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received: it was reported that the new lead was implanted and old product explanted on (b)(6) 2017.Only the open lead wasn't implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key6421496
Report Number2649622-2017-02497
Device Sequence Number474584
Product Code MHY
UDI-Device Identifier00643169313538
UDI-Public00643169313538
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P960009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/06/2019
Device Model Number3389S-40
Device Catalogue Number3389S-40
Device Lot NumberVA1CDNK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2017
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 02/22/2017
Supplement Date Received by ManufacturerNot provided
Not provided
04/27/2017
09/13/2022
Initial Report FDA Received Date03/21/2017
Supplement Report FDA Received Date04/24/2017
04/27/2017
10/02/2017
09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/07/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age34 YR