• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616S
Device Problems Device Operates Differently Than Expected (2913); Temperature Problem (3022); Noise, Audible (3273)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/1901
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the mdu was making a grinding noise and getting hot.A backup device was available to compete the procedure without delay or patient impact.
 
Manufacturer Narrative
Device investigation narrative - one service replacement powermax elite motor drive unit, part number 72200616s was received on (b)(6) 2017 and confirmed to be serial number (b)(4).Complaint of overheating and grinding noise was confirmed.Cause of overheating and grinding noise is a corroded motor/gearbox.The motor/gearbox was removed from the housing.The gearbox was removed from motor and motor tested good.The gearbox was found to be jammed and corroded internally.A review of the manufacturing records shows that this unit was released to distribution on or about (b)(6) 2015.The root cause of this event was identified to be associated with corrosion of the motor and gearbox assembly.A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC REPL,MDU, HAND CNTRL, PWRMX
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key6421662
MDR Text Key70694380
Report Number1643264-2017-00155
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/13/2017
Initial Date FDA Received03/21/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/23/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-