• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number U3513038
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/27/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device was returned to the manufacturer for evaluation.The investigation is currently under way.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the pta balloon ruptured (below nominal pressure) during the first inflation in the distal region of the popliteal artery.The balloon catheter was exchanged over the guidewire for another and the procedure was completed.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only event reported to date for this lot number and failure mode.Visual inspection: the sample was returned with the product label.The balloon size for this product was printed on the balloon hub of the catheter and identified the returned sample as a 3.0 mm x 80 mm balloon.A circumferential balloon rupture was present 0.8 cm from the distal tip.No other anomalies were noted along the length of the catheter.Functional/performance evaluation: functional testing could not be performed due to the condition in which the sample was received.Medical records review: medical records were not provided for review.Image/photo review: images/photos were not provided for review.Conclusion: the device was returned.The investigation was confirmed for a circumferential balloon rupture based on the condition of the returned sample.The definitive root cause could not be determined based upon available information.It was unknown whether patient and/or procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: do not exceed the rated burst pressure (rbp) recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.Precautions: if resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and introducer sheath/guide catheter s a single unit.Use of the ultraverse 035 pta dilatation catheter: position the balloon relative to the lesion to be dilated, ensure the guidewire is in place, and inflate the balloon to the appropriate pressure.Potential adverse reactions: additional intervention the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the pta balloon ruptured (below nominal pressure) during the first inflation in the distal region of the popliteal artery.The balloon catheter was exchanged over the guidewire for another and the procedure was completed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035 PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6423275
Report Number2020394-2017-00234
Device Sequence Number16281684
Product Code LIT
UDI-Device Identifier00801741091698
UDI-Public(01)00801741091698(17)170131(10)50120182
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2017
Device Catalogue NumberU3513038
Device Lot Number50120182
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/23/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/21/2017
Supplement Report FDA Received Date04/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
Patient Weight63
-
-