U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CARBOFLO VASCULAR GRAFT; EPTFE VASCULAR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CARBOFLO VASCULAR GRAFT; EPTFE VASCULAR GRAFT Back to Search Results
Catalog Number F8008TWSC
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device is not available for return.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an intraoperative procedure, upon removal of the beading from the vascular graft, the graft allegedly tore.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and the lot met all release criteria.The dhr review showed that all process parameters were within specification and that this lot of grafts passed all physical testing requirements.There was no evidence of any deficiencies within the lot that would contribute to this complaint.There was nothing found to indicate there was a manufacturing related cause for this event.Visual inspection: the sample was not returned; therefore, a visual inspection could not be performed.Functional/performance evaluation: the sample was not returned; therefore, functional/performance evaluation could not be performed.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: the device was not returned for evaluation.Images and medical records were not provided for review.The investigation was inconclusive for torn material.The definitive root cause could not be determined based upon available information.It was unknown whether patient and/or procedural issues contributed to the event.Labeling review: the current instructions for use (ifu) states: precautions when removing the external spiral support (beading) of the graft, the beading must be removed slowly and at a 90° angle to the graft.Rapid unwinding and /or removal at less than a 90° angle may result in graft damage.Do not use surgical blades or sharp, pointed instruments to remove the beading as this may damage the graft wall.If damage occurs, that segment of the graft should not be used.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an intraoperative procedure, upon removal of the beading from the vascular graft, the graft allegedly tore.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARBOFLO VASCULAR GRAFT
Common Device Name
EPTFE VASCULAR GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6423276
Report Number2020394-2017-00233
Device Sequence Number19027678
Product Code DYF
UDI-Device Identifier00801741021640
UDI-Public(01)00801741021640(17)210627(10)VTAT0739
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K004012
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 04/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/27/2021
Device Catalogue NumberF8008TWSC
Device Lot NumberVTAT0739
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 02/22/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/21/2017
Supplement Report FDA Received Date04/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-