| Model Number |
ZLB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Halo (2227); No Code Available (3191)
|
| Date of Event |
12/02/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to the manufacturer has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zlb00 22.5 diopter intraocular lens (iol) was explanted from a patient's right eye due to patient experiencing very blurred vision with halos, affecting the patient's driving.The lens was replaced with the same model lens with a 21.0 diopter.There was no incision enlargement or vitrectomy required.Additionally, there was no patient injury reported for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation ¿ yes, returned to manufacturer on 3/24/2017.Device returned to manufacturer ¿ yes.Device evaluation the device was returned to the manufacturer.Visual inspection was performed and lens was observed cut in pieces, most probably to make explant possible.Considering the condition of the lens, additional analysis is not possible.The customer's reported complaint cannot be confirmed.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no similar complaints were received for this order number to date.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the device.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
|
| |
|
Search Alerts/Recalls
|