• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C1240
Medical Device Problem Codes Bent (1059); Failure to Fold (1255); Device Handling Problem (3265)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2017-00417, 3005168196-2017-00418, 3005168196-2017-00419, 3005168196-2017-00457.The hospital disposed of the device.
 
Event or Problem Description
The patient was undergoing a coil embolization in the external and internal iliac artery using ruby coils and pod packing coils (podj coils).During the procedure, the physician advanced two ruby coils in the target vessel using a lantern delivery microcatheter (lantern); however, due to a very high flow in the internal iliac artery, both ruby coils would not take their intended shape and anchor in the aneurysm.After several failed attempts, the ruby coils were removed and a pod8 coil was successfully deployed and detached in the aneurysm.The scrub technologist then inadvertently bent both ruby coils pusher assemblies while attempting to re-sheath them.Next, the scrub technologist inadvertently pulled off the introducer sheath of podj coil after removing it from the packaging hoop.While attempting to re-sheath it, the podj coil pusher assembly was accidentally broken; therefore, the podj coil was set aside.The physician then opened a new podj coil and attempted to advance it through the lantern; however, resistance was experienced prior to the coil exiting the lantern and the physician decided to retract it.Upon retraction, the podj coil unintentionally detached from the pusher assembly and unraveled inside the lantern.The lantern was then pulled out of the other manufacturer¿s 5f catheter with the detached coil still inside.It was reported that the technician dropped the first lantern; therefore, it was not used for the remainder of the procedure.The procedure was completed using another lantern and additional ruby coils.In addition, there was no alleged deficiency with the first lantern.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6424673
Report Number3005168196-2017-00420
Device Sequence Number18983305
Product Code HCG
UDI-Device Identifier00814548013015
UDI-Public00814548013015
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date11/15/2023
Device Catalogue NumberRBY2C1240
Device Lot NumberF66530
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2017
Initial Report FDA Received Date03/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-